T90 Below 15 Seconds
Specified for rapid response to changing carbon monoxide concentration.
The MEu-CO is a fuel-cell carbon monoxide sensor with a 0–1000 ppm range and a specified T90 response time below 15 seconds. It is designed for portable instruments, fixed detectors and industrial CO monitoring systems.
Fast electrochemical carbon monoxide sensing with stable repeatability for portable and fixed CO detectors.
The sensor generates a current proportional to carbon monoxide concentration through an electrochemical oxidation-reduction reaction.
Its fast response, low zero output and integrated filter support CO detection in petroleum, petrochemical, metallurgy, mining and environmental applications.
Specified for rapid response to changing carbon monoxide concentration.
A filter helps reduce the influence of acid gases and organic vapors.
Supports concentration calculation through the external signal-conditioning and calibration circuit.
Designed for consistent response across repeated exposure and recovery cycles.
Provides selectivity for CO detection when used according to the datasheet and application guidance.
Available for integration into fixed, portable, residential and industrial gas-detection products.
| Parameter | Specification |
|---|---|
| Model | MEu-CO |
| Detected Gas | Carbon monoxide (CO) |
| Measurement Range | 0–1000 ppm |
| Maximum Range | 2000 ppm |
| Sensitivity | 0.085±0.015 nA/ppm |
| Response Time (T90) | <15 s |
| Zero Output in Clean Air | <±2 ppm equivalent CO |
| Zero Drift (-40°C to 50°C) | <+10 ppm equivalent CO |
| Repeatability | <±3% of output value |
| Linearity | Within ±5% |
| Stability | Average decay <2% per month |
| Filter | Acid gas and organic vapor |
Specifications are based on the official Winsen product page or product manual. Confirm the latest datasheet, calibration requirements, communication protocol and mechanical drawing before final product design.
Contact us for the latest datasheet, model selection support, samples and OEM integration information.