MD62 Thermal Conductivity CO2 Gas Sensor

The MD62 is a thermal-conductivity CO2 gas sensor for industrial carbon-dioxide detection. Its active and reference elements form a Wheatstone bridge, providing a linear concentration-related voltage change with temperature compensation.

BrandWinsen
Sensing PrincipleThermal Conductivity
Target GasCO2
Product TypeThermal Conductivity CO2 Gas Sensor
CO2 Sensing for OEM Integration

Product Overview

A compact thermal-conductivity sensing element for industrial CO2 detection, including oxygen-deficient environments.

The sensor contains matched active and reference elements in a Wheatstone bridge. Changes in the thermal conductivity of the surrounding gas alter the bridge output in relation to gas concentration.

The reference element provides a baseline for temperature compensation, supporting reproducibility and reliable performance in suitable industrial instruments.

Designed for Reliable Integration

Key Features

Thermal Conductivity Principle

Measures gas-composition changes through the thermal response of matched sensing elements.

Linear Output Signal

The product page specifies a linear bridge output related to CO2 concentration.

Good Reproducibility

Designed for repeatable and reliable performance in industrial equipment.

Poisoning Resistance

The sensing structure is described as resistant to toxicosis.

Low-oxygen Operation

Can detect CO2 without oxygen or under oxygen-deficient conditions.

Compact Element

The sensor dimensions are 19 × 9.5 × 14 mm.

Typical Use Cases

Applications

Industrial CO2 Detection
High-concentration Gas Monitoring
Oxygen-deficient Environments
Embedded Gas Analyzers
Official Product Parameters

Technical Specifications

Parameter Specification
Model MD62
Detected Gas Carbon dioxide (CO2)
Detection Principle Thermal conductivity
Working Voltage 3.0±0.1 V DC
Working Current ≤120 mA
Sensitivity ≥5 mV per 10% CO2
Linearity ≤5%
Output Signal Linear Wheatstone-bridge voltage signal
Response Time, 90% ≤15 s
Recovery Time, 90% ≤30 s
Operating Environment -20°C to +50°C; <95% RH
Storage Environment -20°C to +70°C; <95% RH
Dimensions 19 × 9.5 × 14 mm
Explosion-proof Marking Exdib I

Specifications are based on the official Winsen product page. Confirm the latest datasheet, communication protocol, calibration requirements and mechanical drawing before final product design.

Need More Information About the MD62?

Contact us for the latest datasheet, model selection support, samples and OEM integration information.